ChemNet
 
Previous article Next article Contents  

T. A. Filimonova, V. V. Senyuta, M. A. Proskurnin, A. V. Fokin, O. B. Ovchinnikov, S. V. Egerev

Estimation of performance parameters of afrontal variant of optoacoustic determination with a model system

Abstract

Using a model system, the applicability of the frontal configuration of optoacoustic spectroscopy in analytical practice is shown. An optoacoustic cell made using layered-prism geometry was designed and tested; it provides the work with layer thickness of 20–100 µL and sample volumes 50–100 µL. The conditions for the determination of iron(II) as tris(1,10-phenanthrolinate) with the detection limit of 3×10–7 M, which is lower that the detection limit for similar excitation conditions and for the same determination procedure for thermal-lens spectrometry (4×10–7 M). The determination has high reproducibility (relative standard deviation for 10–6M is 0.005) and rapidity (1 min per sample detection).
Moscow University Chemistry Bulletin.
2011, Vol. 52, No. 3, P. 209
   

Copyright (C) Chemistry Dept., Moscow State University, 2002
   Overview
   Editorial board
   Tables of Contents
   Subscription

The site is supported by Russian Foundation for Basic Research
  The using of published on this page materials is not allowed without special permission
Copyright (C) Chemisty Department of Moscow State University
Web-Editor: B.I.Pokrovskii
Web-design: Copyright (C) MIG and VVM
webmaster@www.chem.msu.su